# Safety: hazards, controls, commissioning

This kit puts a mains-powered pulse source next to a corrosive liquid and makes a small
electrical explosion on purpose. The design reduces the risks; it does not remove them, and
nobody has built or tested it yet.

- **It is not certified.** No CE/UKCA/UL assessment, no IEC 61010-1 type test. A school must
  treat it as home-built equipment: its own written risk assessment, an electrical inspection and
  test by a competent person, and whatever its science-safety adviser (CLEAPSS in the UK) requires.
- **Mains wiring is an adult job.** Students can build the 12 V control board, print parts and
  fold copper. They must not wire or open the mains side.
- **Teacher demonstration first.** Run it as a demonstration behind a safety screen. Student
  operation of FIRE, under supervision, is a decision for the school's risk assessment.

## Changes from the original brief for safety, and why

| The brief asked for | What the design has | Reason |
|---|---|---|
| Rectifier or AC divider straight from 110/230 V mains | Safety isolating transformer; nothing on the output side is connected to mains or earth | With a mains-referenced output, one hand on the electrolyte and one on a tap is a mains shock. No divider fixes that. |
| 120 V on the cell | 55 V and 110 V levels, the higher one behind a key | 55 V is already above the 40 V at which Matsumoto reports sparks (Ref. 59). Start there. |
| Lead wire | Aluminium, iron or tin for class use; pure tungsten as an option, prepared by the teacher | An arc on lead makes lead fume and scatters lead particles. Lead belongs in a fume cupboard with an adult, not in a classroom. Aluminium and tin bring their own, smaller hazard: see "Aluminium and tin wire in alkali" below. For tungsten see "Tungsten wire (option)". |
| KOH 1.5 mol/L | KOH for teacher-prepared runs; K2CO3 1.5 mol/L as the class alternative | 1.5 mol/L KOH is corrosive and blinds on eye contact. Matsumoto himself used potassium carbonate in Ref. 59; it is an irritant, not a corrosive. |

## Hazards and the controls built in

| Hazard | Control in the design | What people must still do |
|---|---|---|
| **Electric shock from the output** (55/110 V, several amps) | Isolating transformer: output floats. Output relay connects the sockets only for 1.6 s per shot. Cap interlock: relay cannot close unless the cap is on the glass. Shrouded 4 mm safety sockets and leads. Wire clamp screws recessed; red hood over the wire end. Key switch OFF isolates the transformer secondary. | Never defeat the interlock. Key out before touching the cell. Red and black leads no longer than 1 m, undamaged (the 12 V interlock lead may be up to 1.2 m). |
| **Electric shock from the mains** | Fused, double-pole switched IEC inlet. Class II 12 V module. Earthed metalwork. Plastic case with finger-safe 3 mm vents. Printed walls separate the transformer bay from the control board, and the mains inlet and 12 V module from the isolated-side parts. | Adult build, inspected and tested (below). Use an RCD-protected socket. Unplug before opening. |
| **SCR fails short-circuit** (their usual failure) | The relay is a second, independent switch, driven by a hardware timer, so the worst case is one 1.6 s burst per press. OUTPUT lamp sits before the relay: lit at rest means a fault. Nothing indicates a welded relay contact directly; it shows only as about 15 V at the sockets at rest (test 10). | If OUTPUT is lit when nobody is firing: switch off, do not use. |
| **Controller misfires at switch-on or from noise** | Power-on reset holds the timer for 0.4 s. Pulse timer is held in reset except during the relay window. Both opto LEDs in series. No firmware anywhere. | - |
| **Caustic splash** | Cap covers the vessel. Vent is baffled and faces away from the operator. Only about 100 mL of solution. | Splash-proof goggles (not safety spectacles or ordinary glasses), safety screen, teacher makes up and disposes of solution, eyewash to hand. |
| **Bang, ejected droplets, glass** | Energy per shot is capped at roughly 110 J by ballast and timer in normal operation. If an SCR has failed short, the relay window sets the limit instead: one burst of up to 1.6 s, roughly 220-270 J. The vent is always open. The cap is held by friction only and lifts at well under 0.01 bar. | Safety screen. Inspect the glass for chips and cracks before each session. Soda-lime staining troughs are not rated for thermal shock: one shot, then let it rest. |
| **Hydrogen and oxygen** | At most about 0.7 mL of mixed gas per shot into about 450 mL of vented headspace, far below the flammable limit. An aluminium or tin wire also gives off a little hydrogen while its tip is in the solution, with or without a pulse (below). | Do not block the vent. Do not tape the cap down. Keep the time from putting the cap on to FIRE short, and lift the cap off after the shot. With aluminium, the no-pulse blank matters more: the tip changes even without a pulse. |
| **Hot parts, overheating** | Ballast on an aluminium plate with a 70 °C thermal switch in the interlock chain. 24 s lock-out between shots. | - |
| **Metal fume and dust** | - | No lead, cadmium, beryllium-copper or thoriated tungsten. Ventilated room. Wash hands. |

## Aluminium and tin wire in alkali

Aluminium dissolves in potassium hydroxide solution and gives off hydrogen
(2 Al + 2 KOH + 6 H2O -> 2 K[Al(OH)4] + 3 H2). Potassium carbonate solution is less alkaline
but still attacks it, more slowly. Tin is also attacked by strong alkali, slowly when cold.
Iron is not. This is ordinary textbook chemistry; how fast it goes at the wire tip in our two
electrolytes has not been measured.

- **Why it stays small.** Only the tip of the wire touches the liquid, so very little metal is
  exposed. By calculation, each millimetre of 1 mm aluminium wire that dissolves completely gives
  about 3 mL of hydrogen. Hydrogen's lower flammable limit (4 %) in 450 mL of headspace is about
  18 mL, and the headspace is vented. Keep the vent clear (the hydrogen row above).
- **No delay.** The wire touches the solution from the moment the cap goes on. Keep the time from
  then to FIRE as short as the procedure allows, and write it down. Lift the cap off after the shot.
  Do not leave aluminium or tin wire standing in the solution, and do not drop offcuts into it.
- **The tip changes with no pulse at all.** Aluminium dulls, roughens and may bubble as soon as it
  touches the solution, so the no-pulse (soak) blank matters more for aluminium than for iron:
  hold it in the solution for the same time as a real run, in the same electrolyte.
- **Heat.** The reaction gives out heat. At a wire tip this is negligible; with powder, foil or a
  long immersed length it is not, so do not use aluminium in any other form.
- **To confirm in the stage 1 bench tests:** how fast the tip is attacked in 1.5 mol/L KOH and in
  1.5 mol/L K2CO3, and whether gas at the tip affects contact before the pulse. Until then these
  are expectations, not measurements.

## Tungsten wire (option)

Pure tungsten wire, 1.0 mm, may be used in place of the three class-use wires. It is not supplied in
the kit. The practical source is a pack of pure tungsten welding electrodes, 1.0 x 150 mm, marked WP
(green band), each cut to 116 mm. None of this has been tried on the kit: confirm it in the stage 1 bench tests.

- **Pure tungsten only.** Buy wire or rod sold as pure tungsten (99.9 % or better). If you use welding
  electrodes as the source, only the pure type (marked WP, green band) will do. **Never use thoriated
  electrodes (WT, red, yellow, purple or orange band): they contain radioactive thorium.** Ceriated,
  lanthanated and zirconiated types are not pure tungsten either.
- **Cutting is a teacher's job.** Tungsten is hard and brittle. Side cutters shatter it and throw
  sharp fragments. Notch it with a diamond file or an abrasive wheel and snap it inside a folded
  cloth, wearing eye protection, away from students. The cut end is sharp. Do not try to bend it.
- **In the electrolyte.** Cold alkali attacks tungsten only very slowly with no current, so unlike
  aluminium there is no hurry between fitting the cap and FIRE. While current flows, alkaline
  electrolyte etches tungsten: this is the ordinary way tungsten probe tips are sharpened, so a
  thinned or pointed tip after a shot is expected electrochemistry.
- **Fume and deposits.** An arc on tungsten gives a little tungsten oxide. It is far less hazardous
  than lead fume, but keep the same screen, vent and goggles, and treat deposits as samples, not dust.
- **Disposal.** Used solution from tungsten runs contains tungstate: do not wash it away; use the
  school's metal waste route.

## Numbers behind those claims (estimates, to be measured on the first unit)

- Transformer 80 VA, 2 x 55 V: about 61 V per winding off load. Internal resistance referred to
  the secondary roughly 16.6 ohm (series) or 4.1 ohm (parallel), from the maker's typical regulation of 11 %.
- With the 10 ohm ballast: short-circuit current about 4.6 A at 110 V and 4.3 A at 55 V.
- Most power the cell can ever take: about 140 W (110 V) or 66 W (55 V). Over 0.8 s that is at
  most about 110 J, enough to boil off roughly 0.04 g of water (about 70 mL of steam).
- With a shorted SCR the burst lasts the whole 1.6 s relay window: roughly twice the energy and charge above.
- Charge per shot at most 3.7 C: about 0.5 mL hydrogen plus 0.2 mL oxygen.

## Commissioning tests (do all of them, in this order, and write down the results)

**A. Board alone, on a 12 V bench supply, nothing else connected**
1. Fit a 1 kohm resistor and LED from terminal C to KD in place of the relay coil, and link VCC to C in place of the interlock.
2. Power up. Nothing may flash. POWER and CAP light.
3. Press FIRE. The relay LED lights for about 1.6 s, WAIT for about 24 s. PULSE flashes for the selected time.
4. Press FIRE again during WAIT: nothing happens. Hold FIRE down: still one shot.
5. Remove the VCC-C link. Press FIRE: the relay LED must not light and PULSE must not flash.

**B. Chassis, cold (unplugged)**
6. Earth continuity from the inlet earth pin to the aluminium plate, toroid bolt and switch bodies: under 0.1 ohm.
7. Insulation resistance at 500 V DC: live+neutral to earth, live+neutral to the output sockets, output sockets to earth. Each over 20 Mohm.
8. Check the primary links match your mains voltage, and the label on the case says so.

Not a bench test, but required before any unit leaves the builder: a dielectric-strength test between mains and the isolated output, and between mains and the 12 V side, at the voltage the applicable standard calls for, done by a competent tester as part of the independent assessment.

**C. Powered, with a dummy load instead of the cell**
9. Load: a 100 ohm 50 W resistor across BOX and WIRE, and a shorting plug in the interlock jack.
10. Key OFF: output 0 V. Key at 55 V, at rest: output sockets 0 V, OUTPUT lamp dark. A steady reading of about 15 V at the sockets at rest means a welded relay contact: do not use. Repeat at 110 V, and confirm at both levels that the lamp is dark at rest and lit during a pulse.
11. FIRE at 800 ms: one burst. Confirm on an oscilloscope (isolated probe or battery scope) that it starts and ends at zero crossings and lasts 0.75 to 0.90 s (38 to 45 cycles at 50 Hz, 45 to 54 at 60 Hz).
12. Repeat at 110 V. Measure the off-load voltage at the SCR terminals: expect 115-125 V.
13. Pull the interlock plug: FIRE must give nothing.
14. Short BOX to WIRE and fire once at 110 V: the unit must survive; the ballast gets warm. Fuse F2 may open during this test. That is acceptable: fit a new one and record it.
15. Phase check done at build time (see Electronics_README.md) recorded.

**D. With the cap**
16. Plug the cap in but leave it off the vessel: CAP OK dark, FIRE gives nothing.
17. Seat the cap on the empty vessel: CAP OK lights. Lift one edge 3 mm: it goes out. Adjust the microswitch until it does.

Only then use electrolyte, starting at 55 V and 20 ms.

## Routine checks before each session

Leads and plugs undamaged; glass free of chips; vent clear; CAP OK goes out when the cap is
lifted; OUTPUT lamp dark at rest; key in the teacher's pocket.

## Disposal

Neutralise used KOH or K2CO3 with dilute citric or ethanoic acid to pH 6-9 and wash away with
plenty of water, unless the solution contains anything other than copper, aluminium, iron or tin,
in which case follow the school's heavy-metal waste route. Solution used with aluminium wire may
turn cloudy white as it is neutralised (aluminium hydroxide); that is expected. Keep used wires and boxes dry in labelled bags
for microscopy.